BABY INCUBATOR D-555

Physical Attributes Hood Specifications
Height (Adjustment) 126.1 cm to 146.1 cm Tubing Access Ports 10
Width 99.7 cm Hand Ports / Iris Ports 6
Depth 57.7 cm Hood Type Single / Double / Triple
incubator-Catalogue-13-page-4-final-2022
2
Physical Attributes Hood Specifications
Height (Adjustment) 126.1 cm to 146.1 cm Tubing Access Ports 10
Width 99.7 cm Hand Ports / Iris Ports 6
Depth 57.7 cm Hood Type Single / Double / Triple
Weight 120 kg Soft Bed Mattress Size 76.3 cm x 40.6 cm x 2 cm
Gross Weight 125 kg Baby Bed Inclination 12⁰
Package Dimensions 107 x 68 x 155 cm X-Ray Tray X-Ray Cassette Tray
Temperature Control Modes
Control Type Microprocessor Controlled Unit
Temperature control modes Skin and air temperature servo control mode
Air mode control temperature range 20 ⁰C ῀ 37 ⁰C
Air mode control override temperature range 37 ⁰C ῀ 39 ⁰C
Air mode control accuracy ≤ ± 1.0 ⁰C
Skin mode control temperature range 34 ⁰C ῀ 37 ⁰C
Skin mode control override temperature range 37 ⁰C ῀ 38 ⁰C
Skin temperature accuracy ± 0.3 ⁰C
Dual Skin temperature monitoring Yes
Performance
Air flow velocity across mattress < 10 cm/sec
Temperature rise time at 22⁰C (72⁰F) ambient < 40 min (From 22 ⁰C, 50 % RH, to 35 ⁰C)
Temperature variability < 0.5 ⁰C
Operating noise level in hood ≤ 45 dBA
Carbon dioxide (CO2) Level < 0.5 ⁰C
Test System Self-Test / Silence / Test Button
Micro air intake filter Hepa 13 99.95 %
Trends data 3/24/170 Hrs.
Servo Humidity Control
Humidity control range 30 % ῀ 95% RH
Humidity control operating time without refilling 24 hours, support add water anytime
Humidity control reservoir capacity 1500 ml
Humidity display and control accuracy ± 5% RH
Servo Oxygen Control Option
Oxygen control range 21 % to 65 %
Oxygen display accuracy of full scale ± 2 %
Oxygen control accuracy (100 % calibration) ± 3 %
Oxygen control accuracy (21 % calibration) ± 5 %
Oxygen display resolution 1 %
Oxygen Flowmeter 0-15 liter/min
Oxygen regulation (optional) Oxygen inlet port with flowmeter

Oxygen servo in hood

SPO2 Sensor
Pulse Oximeter (SPO2) Built in pulse oximeter with reusable sensor
Weight Scale Option
Weight range 300 g to 8 kg
Weight display resolution 1 g
Weight accuracy ± 10 g
Operating environment 20 ⁰C to 30 ⁰C
Temperature 10 to 95 % RH
Air Velocity Up to 0.3 m/sec
Storage / Shipping Information
Temperature -25 ⁰C to 60 ⁰C
Humidity 0 to 95 % RH
Safety Alarms
Visual & Audible Alarms High/Low Temperature, Technical fault, Sensor failure Skin Probe

disconnect/damage, Fan/ Ventilation Failure, Mains Failure

Power Requirements
Mains Voltage 220/230 V, +10%, 50Hz